Surface Probe Calibrator
Use our Feedback Form for more information, to request a catalog or for a Price Quote
Surface probes are difficult to calibrate because its hard to find a flat, heated surface thats stable and uniform. Harts new Model 3125 Surface Dry-Well takes advantage of our proprietary Model 2200 Temperature Controller and gives you the best possible conditions for calibrating surface sensors. Why buy a non-temperature calibration device designed for test tube sterilization or PC board repair when you can have a true calibration instrument? The 3125 has a uniform surface temperature and reaches temperatures as high as 400°C. The test surface is milled aluminum for an absolutely smooth and true calibration work area with maximum thermal conductivity. The 12.25-square-inch test surface is large enough to calibrate more than one sensor at a time. The 3125 can be used with a reference surface sensor or PRT. PRTs (3/16" diameter, such as the Model 5613) may be inserted through a drilled hole into the center of the block for use as reference thermometers or for easy recalibration of the units display. With an accuracy of ±0.5°C to 200°C and ±1°C to 400°C, you can calibrate almost any surface probe, thermistor, thin film sensor, RTD, thermocouple, ribbon sensor or surface mount cutouts, fuses, and switches. Stability is within ±0.3°C at 400°C and uniformity within the center three inches of the surface is ±0.6°C at 200°C. Dont buy make-do hot plates when you can have a legitimate calibration tool. Use our Feedback Form for more information, to request a catalog or for a Price Quote
|
||||||||||||||||||||||||||||||||||||||||