Portable Lab Dry-Wells

Portable Lab
Dry-Wells Models 9007 and 9023 |
Designed for on-ship and on-the-go lab
applications |
Two models covering 40°C to
650°C |
Includes three-inch and six-inch
calibration zones |
Use our Feedback Form for more
information, to request a catalog or for a Price Quote
Specifications
|
Range |
9007: 40 to 140°C at
25°C
9023: 35 to 650°C |
Accuracy
|
9007: ±0.15°C
9023: ±0.3°C to 450°C, ±0.5°C to 650°C |
Stability
|
9007: ±0.02°C
9023: ±0.08°C |
Heating
Times |
9007: 25 to 140°C: 20 min.
9023: 25 to 650°C: 45 min. |
Cooling
Times |
9007: 25 to 40°C: 25
min.
9023: 650 to 100°C: 150 min. |
Stabilization
|
10 minutes |
Test
Wells |
9007: 3/4" dia. x
6" deep
9023: 1" dia. x 6" deep |
Communications
|
RS-232 and IEEE-488 incl. |
Enclosure
|
Meets Type II, Class 3, Style D
requirements of MIL-T-28800 |
Power |
9007: 115 VAC (±10%), 5 A
or 230 VAC (±10%), 2.4 A, switchable, 50/60 Hz, 560 W
9023: 115 VAC (±10%), 7 A or 230 VAC (±10%), 3.4 A, specify, 50/60 Hz, 800 W
|
Size |
13.8" H x 10.8" W x
16.9" D (35.1 x 27.4 x 42.9 cm) |
Weight
|
36 lbs. (16.3 kg) |
NIST-Traceable
Calibration |
9007: Data at
40°C, 0°C, 25°C, 75°C, 140°C;
9023: Data at 50°C, 100°C, 200°C, 300°C, 400°C, 500°C, 650°C |
These calibrators cover temperatures from 40°C to 650°C, deliver
the performance youll only find in true lab standards, and come in a totally
portable case. If your work involves ocean vessels, aircraft, or service trucks, these
dry-wells were designed for you. The U.S. Navy purchases the Model 9007 on contract
N00244-00-D-0100.
The Model 9007 Dry-Well covers 40°C to 140°C. No external cooling
is needed, so you get 40°C in normal ambient. Set-point accuracy is ±0.15°C and
stability is better than ±0.02°C.
For hotter temperatures, the Model 9023 Dry-Well extends from 35°C to
650°C. Heating time is only 45 minutes to 650°C. Set-point accuracy is ±0.3°C to
450°C and ±0.5°C to 650°C. Stability is better than ±0.08°C even at 600°C.
These dry-wells come with a unique NIST-traceable calibration at the full,
six-inch depth of the well and at a three-inch depth for short probes. Coefficients for
each calibration are stored in the dry-well and can be easily selected from the
front-panel control buttons to match the length of the probe being tested.
Both models are encased in all-aluminum enclosures that are durable,
waterproof, and meet the standards of MIL-T-28800. Each unit comes with both RS-232 and
IEEE-488 interface connections. A wide variety of inserts are available covering probe
widths from 1/16" (1.6 mm) to 7/8" (22 mm).
Use our Feedback Form
for more information, to request a catalog or for a Price Quote
|